AESCULAP AG AS VEGA PS FEMORAL COMP.CEMENTED F5N L; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NX011Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Date of Event 10/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reference code (b)(4), device name as vega ps tibial plateau, cemented t2, serial number n/a, batch number unknown, udi device identifier (b)(4), udi production identifier unknown, basic udi-di n/a, unit of use udi-(b)(4), manufacturing date unknown.Ref.Code, device name batch: nx011z, as vega ps femoral comp.Cemented f5n l, unknown; nn260p, plug f/tibial plateau, 51860099; nx044, patella 3-pegs p4, unknown; nx120, vega ps gliding surface t2/2+ 10mm, unknown.Investigation: no product at hand.Therefore an investigation at the product is not possible.Pictorial documentation: there are no pictures available.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Explanation and rationale: in the light of the little information received and due the circumstance that we did not received the complained devices it is not possible to determine a root cause for the mentioned failure.Corrective action: for this topic (vega loosening) a product safety case (psc) was initiated.
 
Event or Problem Description
It was reported that there was an issue with as vega knee.As a result of having the product implanted the patient has experienced knee pain, swelling, difficulty walking, and loosening and instability of the implant.The primary surgery occurred on (b)(6) 2012 and the revision surgery occurred on (b)(6) 2016.All available information has been provided at this time.If additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference (b)(4).Associated medwatches: 2916714-2020-00407, 2916714-2020-00408, 2916714-2020-00409, 2916714-2020-00410.Involved components: nx011z (ps femur cemented f5n lt), nn260p (peek plug f/ tibia), nx053z (ps tibia cemented t2), nx044 (universal patella p4), nx120 (ps pe insert t2/t2+, 10mm).The cement used is unidentified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS FEMORAL COMP.CEMENTED F5N L
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10635830
Report Number2916714-2020-00406
Device Sequence Number19661111
Product Code HRY
UDI-Device Identifier04046963809904
UDI-Public4046963809904
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX011Z
Device Catalogue NumberNX011Z
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/15/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-