| Model Number |
X SERIES |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device inappropriately shut down.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.The device was put through extensive testing including bench handling and full functionality stress testing with a known good ecg cable and multifunction cable on a simulator.Without duplicating the report.Multifunction cable and receptacle were replaced as a precaution.The device successfully passed the final test procedure, was recertified, and returned to the customer.The multi-function cable used at the time of the reported event was not returned for evaluation.Analysis for reports of this type has not identified an increase in trend.
|
| |
|