| Model Number |
D013 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Telemetry Discrepancy (1629); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Electric Shock (2554); No Code Available (3191)
|
| Date of Event |
08/31/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) was unable to be interrogated during a routine follow up.This device was also reported to have emitted beeping tones.A changeout procedure was scheduled.Prior to change out, the patient reported receiving nine shocks due to atrial fibrillation (af).This device was then explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient code 3191 was applied to capture the reportable event of surgery.The product has been received for analysis.This report will be updated upon completion of analysis.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) was unable to be interrogated during a routine follow up.This device was also reported to have emitted beeping tones.A changeout procedure was scheduled.Prior to change out, the patient reported receiving nine shocks due to atrial fibrillation (af).This device was then explanted and replaced.No additional adverse patient effects were reported.
|
| |
|