It was reported that prior to the implant procedure, the physician observed rust-like material near the device header.The physician elected to exchange the device to resolve the event and no adverse consequences were reported.The device is expected for analysis, but has not yet been received.
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header confirmed the presence of unpolished metal and no further anomalies were noted.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|