• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
The device used for treatment was returned for evaluation.Nothing was identified visually that contributed to the problem.A functional evaluation was performed and the reported problem was confirmed.The handpiece connector test failed, when flexing the cable/connector created an over current error.A short was found between the phase and ground wires, indicating failure.The cord was dis-assembled at strain relief, where it was found that excessive epoxy had leaked down inside the cord and most likely is the problem.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints associated with the handpiece and failure mode(s) "error message(s)" identified similar events.The most likely cause of this event is the electrical failure of the cable due to the epoxy leakage found within the cable cord.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that during the tka exposure control motor failed.Back-up device was available.Delay of less than 30 minutes and no patient injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
Manufacturer Contact
rick confer
2828 liberty ave
suite 100
pittsburgh, PA 38116
4126833844
MDR Report Key10638387
MDR Text Key210189407
Report Number3010266064-2020-01817
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-