The device was returned and analyzed.Based on the damage found on the lead, the most likely cause was that the lead caught on a sharp object, such as the needle tip, but the cause was unable to be confirmed.The manufacturing records were reviewed and no issues were found.
|
It was reported to nevro that during the implant procedure, the lead was damaged.The lead was replaced and the procedure completed with no further complications.The device was returned for analysis, which showed an electrode was detached.The patient is currently using their device to find effective pain relief.
|