MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; LASER THERAPY PRODUCT
|
Back to Search Results |
|
Model Number PED2-400-14 |
Device Problems
Break (1069); Material Deformation (2976)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received a report that the pipeline was placed in the desired position, and then deployment was started.However, the sleeves were not opened, and the physician decided to resheath he device.During resheathing, the pipeline left its delivery wire and moved to the distal coil segment.The physician pulled back the system inside the navien catheter.Another pipeline device was successfully implanted and confirmed via angiography post-procedure.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.The patient was undergoing surgery for treatment of a saccular, ruptured aneurysm of the right internal carotid artery with a max diameter of 5.4 mm and a 3.8 mm neck diameter.It was noted the patient's vessel tortuosity was minimal.Dual antiplatelet therapy (dapt) was administered: plavix 6 tablets before three days and continue 1x1 until procedure.
|
|
Event Description
|
No additional information received.
|
|
Manufacturer Narrative
|
H3: analysis of the pipeline flex embolization device (lot no.B025204) found that the pipeline flex pusher was found protruding from within the phenom 27 catheter hub for 42.4 cm.No bends or kinks were found with the phenom 27 catheter body.No damage was found with the phenom 27 distal marker/tip.The pipeline flex ptfe sleeves and tip coil were found protruding from within the phenom 27 distal tip.The ptfe sleeves and tip coil were found in good condition.The phenom 27 catheter total length was measured to be 158.7 cm and the useable length was measured to be 152.3 cm.It could not be determined if the phenom 27 lengths met specification as the model or lot number was not reported.The pipeline flex pusher was pushed out from within the phenom 27 catheter without issue.The pipeline flex pusher, proximal bumper, resheathing pad and resheathing marker were found intact.The pipeline flex braid ends were found open, but slightly damaged (frayed).No other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿pushwire break/separation¿ could not be confirmed.No breaks or separations were found with the returned pipeline flex embolization device.H6: device code updated to a0406.H10: review of analysis results shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned in a way that may have led to a death or serious injury.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received reported that the pushwire separation occurred prematurely.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|