Model Number SONATA |
Device Problem
Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 09/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
It was reported that the user's hearing with the device has degraded in the last 15 days.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode, as it might be caused by an external mechanical impact appears likely.Additionally, in situ measurements show the most basal channel with hi status since 2018 due to undetermined reasons.However to confirm an exact root cause of failure a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
|
Event Description
|
It was reported that the user's hearing with the device has degraded in the last 15 days.
|
|