MEDTRONIC IRELAND ENDURANT II EXTENSION CUFF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETTF2828C70E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An aneurx stent graft system (bifurcate and 4 limbs) were implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm.An endurant cuff was implanted approximately 7 years post the index procedure in the then 120mm aaa for the treatment of a type ia endoleak.It was reported approximately 10 years post the index procedure, follow up ct demonstrated a type iiia endoleak where the endurant cuff and aneurx main body separated.Intervention was completed where a valiant navion stent graft was implanted and the endoleak confirmed as resolved.Per the physician the cause of the type iiia endoleak was due to tortuous anatomy and angled neck.No additional sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II EXTENSION CUFF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
IE 
MDR Report Key10641122
Report Number9612164-2020-03799
Device Sequence Number11674789
Product Code MIH
UDI-Device Identifier00613994991492
UDI-Public00613994991492
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2019
Device Model NumberETTF2828C70E
Device Catalogue NumberETTF2828C70E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2020
Initial Report FDA Received Date10/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
-
-