| Model Number |
7266V |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing record was reviewed, there were no nonconformance related to this lot, therefore, supporting the device met material, assembly and performance specifications.A follow-up report will be issued after the investigation is complete.
|
| |
|
Event or Problem Description
|
|
As reported: upon hand injection, sheath separated from hub leaving sheath in the body.Removed with cutdown to remove sheath.Patient is reported as doing well post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
One sheath from model 7266v was returned to vsi for evaluation.The shaft of the sheath was separated about 1-2mm from the hub.The separation point was stretched and torn.The separation was not related to seat depth of the shaft inside the hub.The damage to the shaft was at the separation was consistent with the sheath being pulled against resistance.This is evident by the stretched material.It is unknown if there was damage present to the shaft prior to injection.
|
| |
|