• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number 4252519-02
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive the device involved in the reported event are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: noted a particulate in the flashback chamber, after the iv was started.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).One (1) used sample was received for evaluation.When the sample was visually inspected, a particle was observed on the outside of the flashback chamber.This particulate was removed and sent for ftir analysis.The results of the testing did not match any items in use in the production area.Review of the device history record performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional information becomes available, a follow up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key10643473
Report Number9610825-2020-00228
Device Sequence Number16753918
Product Code FOZ
UDI-Device Identifier04046963165949
UDI-Public(01)04046963165949
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4252519-02
Device Catalogue Number4252519-02
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 09/25/2020
Supplement Date Received by Manufacturer09/25/2020
Initial Report FDA Received Date10/07/2020
Supplement Report FDA Received Date01/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-