• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFCSIGMA POST FEMAUG TRL4MM S3; KNEE INSTRUMENT : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFCSIGMA POST FEMAUG TRL4MM S3; KNEE INSTRUMENT : FEMORAL TRIALS Back to Search Results
Model Number 96-1866
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Last week, rep.Notified me that in some revisions two instruments were damaged and one of them apparently was thrown away.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Investigation summary: the device associated with this report was not returned.The investigation could not verify or identify any product contribution to the current reported event without the device to examine.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIGMA POST FEMAUG TRL4MM S3
Common Device Name
KNEE INSTRUMENT : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10643638
Report Number1818910-2020-21729
Device Sequence Number3158726
Product Code HWT
UDI-Device Identifier10603295240334
UDI-Public10603295240334
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-1866
Device Catalogue Number961866
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/14/2020
Supplement Date Received by Manufacturer10/14/2020
12/08/2020
Initial Report FDA Received Date10/07/2020
Supplement Report FDA Received Date10/19/2020
12/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-