Device evaluated by mfr.: synergy ous mr 2.50 x 20 mm stent delivery system was returned for analysis without the stent.A review of the manufacturing stent profile data was performed and the stent outer diameter was within max crimped stent profile measurement.The balloon cones were reviewed, and signs of positive pressure applied to the cones were noted.Crimp markings are visible on the exposed balloon wall.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section found a 2 mm long tear in the shaft polymer extrusion located at the guidewire exchange port.No other issues were identified during the product analysis.
|