Model Number CI-1600-04
Device Problems Malposition of Device (2616); Impedance Problem (2950); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Head Injury (1879); Insufficient Information (4580)
Event Date 09/08/2020
Event Type  Injury  
Event Description
The recipient reportedly experienced impedance issues due to incorrect electrode position.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top and bottom covers, slices along the lead and the electrode was severed prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed cut electrode wires along the lead.These are believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top and bottom covers, slices along the lead, and the electrode was severed near the array prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed cut electrode wires along the lead.These are believed to have occurred during revision surgery.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10645418
MDR Text Key210359428
Report Number3006556115-2020-01105
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)180124(17)210131
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2021
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age22 YR
-
-