MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL INTRFX ADV - LG 23MM INSERTER; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL INTRFX ADV - LG 23MM INSERTER; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR Back to Search Results
Model Number 254832
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that before an anterior cruciate ligament reconstruction the large 23 mm sheath inserter was clogged.Another device was used to complete the procedure.No patient consequences or surgical delay reported.The device is available to be returned for evaluation.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> according to the information provided, it was reported by the sales rep via phone that before an anterior cruciate ligament reconstruction the large 23 mm sheath inserter was clogged.The complaint device was received and evaluated.Visual observations confirm that the sheath inserter was found blocked.In addition, the device presented light nicks and marks of wear in the handle.The complaint can be confirmed.The possible root cause for the reported failure can be occurred during cleaning/ sterilization procedures.However, this cannot be conclusive determined.A manufacturing record evaluation was performed for the finished device lot number:1807001, and no non-conformances were identified.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRFX ADV - LG 23MM INSERTER
Common Device Name
ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10645587
Report Number1221934-2020-02907
Device Sequence Number3649220
Product Code LXH
UDI-Device Identifier10886705026302
UDI-Public10886705026302
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number254832
Device Catalogue Number254832
Device Lot Number1807001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Initial Date Received by Manufacturer 09/22/2020
Supplement Date Received by Manufacturer11/03/2020
Initial Report FDA Received Date10/07/2020
Supplement Report FDA Received Date11/04/2020
Patient Sequence Number1
-
-