Model Number V60 |
Device Problem
Defective Alarm (1014)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 07oct2020.
|
|
Event Description
|
The customer called into technical support (ts) reporting that the device stopped working and displayed several alerts.¿blower stalled, auxiliary alarm supply failed, 35 v supply failed, backup alarm failed and ventilator restarted due to anomalies detected during operation.¿ the device was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or user.The patient was placed on another vent when the reported issue occurred.
|
|
Manufacturer Narrative
|
G4:15feb2021; b4:17feb2021.Multiple attempts were made to follow-up with the customer to obtain additional information; however, there was no response.If additional information is received at a later date, this complaint will be re-opened and a supplemental report will be submitted.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|