Model Number 71365703
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/08/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the inspection before a thr is was noticed that the offset broach handle is not locking correctly.Procedure finished with s&n backup device.No surgical delay.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device was manufactured in 2014 and shows signs of extensive use.A functional evaluation of the returned device could not confirm the stated failure mode.The device functioned as intended with a mating part.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OFFSET BROACH HANDLE
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10646842
Report Number1020279-2020-05232
Device Sequence Number2503941
Product Code JDG
UDI-Device Identifier03596010553393
UDI-Public03596010553393
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71365703
Device Catalogue Number71365703
Device Lot Number14LM14094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Initial Date Received by Manufacturer 09/09/2020
Supplement Date Received by Manufacturer12/03/2020
Initial Report FDA Received Date10/08/2020
Supplement Report FDA Received Date12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-