| Model Number |
PRD-SC30-200 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Rupture (2208)
|
| Date of Event |
09/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.(b)(4).
|
| |
|
Event or Problem Description
|
|
Orbital atherectomy(oa) followed by balloon angioplasty was successfully completed in a calcified superficial femoral artery.Oa treatment was continued in a calcified profunda artery.On the third treatment pass on high speed the stealth peripheral orbital atherectomy device (oad) stopped spinning and the crown was angled outside of the vessel.An angiogram was performed and revealed a vessel rupture had occurred during treatment and that an aneurysm was present near the crown of the oad.The oad and viperwire guide wire were removed.Coiling was performed to resolve the aneurysm.The remaining vessels of the profunda artery were intact, with no other complications observed.The patient condition was reported as fine.Per the opinion of the physician, the csi device contributed to the aneurysm event.The cause of the vessel rupture was unknown.
|
| |
|