The investigation has determined that a lower than expected calcium result was obtained from a non-vitros biorad quality control (qc) fluid using vitros chemistry products ca slides lot 0345-0601-1498 on a vitros 5600 integrated system.The assignable cause of the event is due to a suboptimal calibration.The cause of the suboptimal calibration is unknown.The result was obtained on a calibration that was determined to have atypical parameters when compared to expected database values.When the calibration with acceptable parameters was restored, acceptable results were obtained for the biorad qc and patient sample repeat results.The customer confirmed that they are following the instructions for use for calibration preparation, therefore improper protocol is not a likely cause of the suboptimal calibration.
|