• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. LINEAR HIP; LINEAR, LATERAL, REDUCED NECK, SZ 11 W/P2 COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. LINEAR HIP; LINEAR, LATERAL, REDUCED NECK, SZ 11 W/P2 COATING Back to Search Results
Model Number 425-99-011
Health Effect - Clinical Code Fall (1848)
Date of Event 09/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient fell and broke femur.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as broken femur after a fall.The previous surgery and the surgery detailed in this event occurred 5 months and 3 weeks apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to broken femur after a fall.There were no findings during this evaluation that indicate the reported devices were defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Agent has clearly mentioned that "patient fell and broke femur", it seems that the event may possibly occurred due to poor bone density, patient bone deterioration, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR HIP
Common Device Name
LINEAR, LATERAL, REDUCED NECK, SZ 11 W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10648931
Report Number1644408-2020-00890
Device Sequence Number16884890
Product Code LPH
UDI-Device Identifier00888912078979
UDI-Public(01)00888912078979
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120241
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2025
Device Model Number425-99-011
Device Catalogue Number425-99-011
Device Lot Number576G1119
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/10/2020
Supplement Date Received by Manufacturer10/19/2020
Initial Report FDA Received Date10/08/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-323 LOT 862B1163; 400-03-323 LOT 862B1163
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-