MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71335558
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Codes Injury (2348); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during surgery, after inserting the acetabulum cup 58mm, a 58mm poly liner compatible with the 58mm cup was tried to be used but could not be inserted.A second liner was tried (58mm 20 degree elevation type), but the same problem occurred.Finally, additional reaming was performed and a 60mm cup was used to complete the procedure.A delay greater than 30 minutes occurred without significant effect on the patient.
 
Event or Problem Description
It was reported that during surgery, after inserting the acetabulum cup 58mm, a 58mm poly liner compatible with the 58mm cup was tried to be used but could not be inserted, a second liner was tried (58mm 20 degree elevation type) but would still not work and a third liner was tried but the same problem occurred.Finally additional reaming was performed and a 60mm cup was used to complete the procedure.A delay greater de 30 minutes occurred without significant effect on the patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection could not confirm the stated failure.The device exhibits normal wear and use from attempted insertion of the implant.Dimensional evaluation was completed on the returned shell device for all ctq features critical to liner mating and function.No deviations were noted during dimensional evaluation.The failure mode cannot be confirmed via dimensional analysis of the device.A medical investigation was conducted and confirms per complaint the details, we are currently unable to rule out a procedural variance as a contributing factor to the reported event which does not represent a device malfunction.Without the requested clinical information a thorough medical investigation cannot be rendered.According to the report, the surgeon changed 1 shell to a larger size after three different liners would not seat.There were no deviations noted on the dimensional evaluation of the returned shell.Based on the information provided, the surgeon reamed an additional 2mm and a 60mm cup was used to complete the procedure.It was reported there was a delay of > 30-minutes without a significant impact on the patient.Therefore, no further clinical/medical assessment is warranted at this time.Should any additional medical information be provided, this complaint would be re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 58MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10649442
Report Number1020279-2020-05251
Device Sequence Number2605736
Product Code MBL
UDI-Device Identifier03596010598271
UDI-Public03596010598271
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71335558
Device Catalogue Number71335558
Device Lot Number19KW07210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Initial Date Received by Manufacturer 09/17/2020
Supplement Date Received by Manufacturer10/13/2020
09/07/2021
Initial Report FDA Received Date10/08/2020
Supplement Report FDA Received Date10/15/2020
09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71331758 / 15CT52436; 71332758 / 18MM01522; 71335758 / 17KM00158
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-