• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JII UNI TIB BASE SZ 2 LM/RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JII UNI TIB BASE SZ 2 LM/RL; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71935194
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/11/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tka surgery, two inserts did not fit into the jii uni tibial baseplate sz 2 lm/rl.The surgeon then decided to explant the plate and insert a device from a competitor brand.The surgery was successfully completed.It is unknown if this caused surgery to delay.The patient outcome is unknown.
 
Additional Manufacturer Narrative
Results of investigation: it was reported that during surgery, two poly inserts were tested but would not fit the tibial baseplate.Plate was explanted and a competitors device was implanted which caused a posterior bone stock loss.Procedure was concluded without an injury or a delay.It was reported that during surgery, two poly inserts were tested but would not fit the tibial baseplate.Plate was explanted and a competitors device was implanted which caused a posterior bone stock loss.Procedure was concluded without an injury or a delay.The associated device, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of manufacturing records did not reveal any deviation from the standard manufacturing processes.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Based on this investigation, the need for corrective action is not indicated.Some potential causes could include but are not limited to size of device used or user/procedural variance.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JII UNI TIB BASE SZ 2 LM/RL
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10651652
Report Number1020279-2020-05258
Device Sequence Number14134927
Product Code HSX
UDI-Device Identifier00885556578735
UDI-Public00885556578735
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K152315
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71935194
Device Catalogue Number71935194
Device Lot Number19EAP0202
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/22/2020
Supplement Date Received by Manufacturer01/20/2021
Initial Report FDA Received Date10/08/2020
Supplement Report FDA Received Date02/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CN: 71935205, LN 19GAP0117B (X2); CN: 71935205, LN 19GAP0117B (X2)
Outcome Attributed to Adverse Event Required Intervention;
-
-