• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T2+; KNEE ENDOPROSTHESES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T2+; KNEE ENDOPROSTHESES Back to Search Results
Model Number NX054Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Date of Event 10/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reference code: nx011z.Device name as vega ps femoral comp.Cemented f5n l; serial number: n/a; batch number: 51838609; udi device identifier: (b)(4); udi production identifier: (b)(4).Basic udi: di n/a.Unit of use udi: di (b)(4).Manufacturing date 2012-03-30.Ref.Code; device name; batch#: nx04); patella 3-pegs p4 51822655; nx120; vega ps gliding surface t2/2+ 10mm 51875870; nx011z); as vega ps femoral comp.Cemented f5n l 51838609; nn260p; plug f/tibial plateau 51887768.Investigation: no product at hand.Therefore, an investigation at the product is not possible.Pictorial documentation there are no pictures available.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Explanation and rationale: in the light of the little information received, and due the circumstance that we did not received the complained devices, it is not possible to determine a root cause for the mentioned failure.Corrective action: for this topic (vega loosening) a product safety case (psc) was initiated.
 
Event or Problem Description
It was reported that there was an issue with a vega knee components.As a result of having the product implanted the patient has experienced knee pain, swelling, difficulty walking, and loosening and instability of the implant.The primary surgery occurred on (b)(6) 2013, and there was no reported revision surgery.All available information has been provided at this time, if additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference (b)(4).Involved components: nx054z; (ps tibia cemented t2+); nx044; (universal patella p4); nx120; (ps pe insert t2/t2+, 10mm); nx011zp; (ps femur cemented f5n lt); nn260p; (peek plug f/tibia).Associated medwatches: 2916714-2020-00461; 2916714-2020-00462; 2916714-2020-00463; 2916714-2020-00464.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS TIBIAL PLATEAU CEMENTED T2+
Common Device Name
KNEE ENDOPROSTHESES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10652456
Report Number2916714-2020-00460
Device Sequence Number11503937
Product Code JWH
UDI-Device Identifier04046963810436
UDI-Public4046963810436
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2017
Device Model NumberNX054Z
Device Catalogue NumberNX054Z
Device Lot Number51844943
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/15/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-