| Model Number |
37-1173 |
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The manufacturing lot code on the device indicates that the device was manufactured in 2005.Upon evaluation of the returned instrument, it could not be confirmed that there was any aspect out of conformance.Based on the evalution of the device, it can be determined that the injury that occurred was not the result of a malfunction of the device.There has been a total of 195 sold since 2012 with no additional complaints recorded for this product.
|
| |
|
Event or Problem Description
|
|
After the initial procedure, it was noticed there was a site of bleeding in the descending aorta which was complicated further by aortic injury requiring placement of an interposition graft.There was continued bleeding from all the suture sites.Therefore a covered stent was placed in the aorta.
|
| |
|
Additional Manufacturer Narrative
|
|
On 10/8/2020, additional information was provided by the end user that was not initially provided.The end user stated that the device scissorred when clamping the aorta, therefore damaging it during the initial procedure.Our evaluation of the product did not confirm this issue.The product functioned as intended with normal use.The device would only scissor with excessive force applied while torquing the device and purposefully making it scissor.As a result, there is no evidence that there are any defects in material or workmanship of the device.Therefore, no additional actions are necessary.
|
| |
|