It was reported via complaint submission tool that during a shoulder repair, the surgeon was using the penetrating grasper straight to pass suture through the rotator cuff.He tried to open the jaws, heard a pop, and the jaws would not open.Upon investigation, the jaws would not open any longer.The instrument was replaced with a readily available replacement and the case was successfully completed without patient harm or surgical delay.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported via complaint submission tool that during a shoulder repair, the surgeon was using the penetrating grasper straight to pass suture through the rotator cuff.He tried to open the jaws, heard a pop, and the jaws would not open.Upon investigation, the jaws would not open any longer.The instrument was replaced with a readily available replacement and the case was successfully completed without patient harm or surgical delay.The device was received and inspected.Visual inspection reveals that the device lot number 17n02, is worn due to it has marks of used.The device was tested, as a result, the jaws do not open or close.Therefore, the mechanism to close and open the jaws was defective.This complaint can be confirmed.The possible root cause for the reported failure could be related to fair wear and tear.However, this cannot be conclusive determined.A manufacturing record evaluation was performed for the finished device lot number:17n02, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|