MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA ENGINE CANISTER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA ENGINE CANISTER; NRY Back to Search Results
Model Number PAPS3-A
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the internal carotid artery (ica) using a penumbra engine canister (canister), a penumbra engine (engine), a penumbra system ace 68 reperfusion catheter (ace68), a balloon guide catheter and a revascularization device.During the procedure, the devices were set up and connected to the patient and the engine was powered on.The engine indicator lights did not illuminate until pressure was applied by hand to the canister.When the pressure was removed, the lights ceased to illuminate.Therefore, the canister was removed.The procedure was completed using a new canister, the same engine, and the same devices.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA ENGINE CANISTER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10653246
Report Number3005168196-2020-01671
Device Sequence Number11730319
Product Code NRY
UDI-Device Identifier00814548010007
UDI-Public00814548010007
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K180008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPAPS3-A
Device Catalogue NumberPAPS3
Device Lot NumberS11503
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2020
Initial Report FDA Received Date10/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1