It was reported that the stent was damaged post-deployment, fractured, and left unretrieved device fragments requiring additional intervention.The target lesion was located in the distal left anterior descending artery (lad).A 2.25x38mm synergy ii drug-eluting stent was deployed in the lad over a non-bsc guidewire.A 3x20mm synergy was then inserted over the same non-bsc guidewire in the lad but failed to cross through the deployed stent.The 3x20mm undeployed stent was removed and a 3x15 emerge balloon was then advanced for stent post dilation.Another non-bsc guidewire was passed down the lad through the deployed stent as a buddy wire.Another attempt to cross through the previous deployed stent with the 3x20mm synergy stent once again failed.The two non-bsc guidewires were then removed with resistance and found to be damaged.Also, a part of the deployed stent was entangled in them.The stent had fractured and some of the stent remained behind in the lad.A new non-bsc guidewire was then inserted through the deployed stent and a 2.5x30mm non-bsc balloon was used for dilation in the lesion/stent area, then a 2x30mm emerge balloon was used.Two non-bsc stents were then deployed in the distal lad, crushing the remaining stent.Post-procedure echocardiogram was negative and the patient was transferred from cath lab pain free.
|