MEDOS INTERNATIONAL SàRL CAMERA HEAD AC - C-MOUNT; ENDOSCOPIC VIDEO CAMERA Back to Search Results
Model Number 242401
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem Not Applicable (3189)
Event Date 09/23/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Udi: (b)(4).
 
Event Description
It was reported by the sales rep via email that during a shoulder scope the camera head ac - c-mount had a short in the cord.Another device was used to complete the procedure.No patient consequences.2 minutes delay reported.The device is available to be returned for evaluation.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the device was received and evaluated at the service center.The reported complaint that the unit had a short in the cord was unable to be confirmed.However, it was found that the camera head was physically damaged.The repair of the device was however declined and was placed into long term hold.The physical damage to the camera head is most likely a result of user mishandling, or a probable fall.However, since the reported condition is not confirmed, the root cause for the reported failure cannot be determined.A manufacturing record evaluation was performed for the finished device [serial number : (b)(6)], and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: a manufacturing record evaluation was performed for the finished device [1802ce0337] number, and no non-conformances were identified.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAMERA HEAD AC - C-MOUNT
Type of Device
ENDOSCOPIC VIDEO CAMERA
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10653425
MDR Text Key210658139
Report Number1221934-2020-02934
Device Sequence Number1
Product Code FWF
UDI-Device Identifier10886705028733
UDI-Public10886705028733
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number242401
Device Catalogue Number242401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2020
Initial Date Manufacturer Received 09/23/2020
Initial Date FDA Received10/08/2020
Supplement Dates Manufacturer Received10/29/2020
10/30/2020
Supplement Dates FDA Received10/29/2020
10/30/2020
Patient Sequence Number1
-
-