| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump passed the displacement test but motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to motor error alarm.Motor passed motor test.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a motor error.Customer stated drive support cap appeared normal-slightly indented.Customer stated they were able to successfully clear the alarm and were able to rewind.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|