MAUDE Adverse Event Report: TX027 TX-JACKSONVILLE-MCKNIGHT LINER SUCTION 1500CC; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TX027 TX-JACKSONVILLE-MCKNIGHT LINER SUCTION 1500CC; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number 65651-920C
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No samples, photos or lot number were provided by the customer.Therefore, the device history record could not be reviewed.Discussions were held with engineering, quality and production concerning the assignable cause for the complaint "liner doesn't fit properly into outer canister, liner deflates on itself".With the additional information of "the end user indicates using a flex liner inside a crd canister", the team concluded that the wrong liner was placed in the canister.Based on the information provided, the assignable cause is error on the part of the end user.No corrective action is planned.End user used a flex liner in crd canister.The customer will be provided the correct liner number.
 
Event or Problem Description
Based on information received from the customer, the liner didn't fit properly into outer canister.Reportedly the liner deflates on itself.The customer stated they were able to move forward with the procedure and there was no injury to patient.Follow up conversation with end user indicates incorrect liner may have been used (flex liner inside crd outer canister).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER SUCTION 1500CC
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
TX027 TX-JACKSONVILLE-MCKNIGHT
200 mcknight st.
jacksonville TX 75766
MDR Report Key10655043
Report Number1423537-2020-00541
Device Sequence Number3281326
Product Code GCX
UDI-Device Identifier00885380029083
UDI-Public00885380029083
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number65651-920C
Device Catalogue Number65651-920C
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2020
Initial Report FDA Received Date10/09/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-