ABBOTT DIABETES CARE INC FREESTYLE FREEDOM; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 11004-20
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The returned meter was investigated with retained test strips.Meter powered on with button depression and test strip insertion.No new issues were observed.Blank screen was not observed.The complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE FREEDOM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10655223
Report Number2954323-2020-06979
Device Sequence Number3578369
Product Code NBW
UDI-Device Identifier00699073110046
UDI-Public00699073110046
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K092638
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number11004-20
Device Catalogue Number11004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Initial Date Received by Manufacturer 10/07/2020
Supplement Date Received by Manufacturer10/31/2020
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight122
-
-