Model Number 71365801
Medical Device Problem Codes Material Disintegration (1177); Naturally Worn (2988)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during the procedure, thr, that the coating on trial necks wear off, causing that the magnet detaches from the trial neck.The procedure was completed without delay, using the same device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Updated d4: lot number and exp / mnf date.
 
Additional Manufacturer Narrative
H10: h6: the associated device, used in treatment, were returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device's coating is worn off, causing magnet mechanism to malfunction.The device was manufactured in 2013 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY HO TR NECK SZ 1-6
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10655466
Report Number1020279-2020-05289
Device Sequence Number2631389
Product Code JDG
UDI-Device Identifier03596010559876
UDI-Public03596010559876
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71365801
Device Catalogue Number71365801
Device Lot Number13BM02880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Initial Date Received by Manufacturer 09/14/2020
Supplement Date Received by Manufacturer09/14/2020
01/11/2021
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date11/18/2020
01/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-