• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPYSCOPE DS; CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00546600
Medical Device Problem Codes Optical Problem (3001); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.(initial reporter address 1): (b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
Note: this report pertains to a spyglass ds controller and two spyscope ds used during the same procedure.It was reported to boston scientific corporation that a spyglass ds controller and two spyscope ds were used during an endoscopic retrograde cholangiopancreatography (ercp) procedure in the common bile duct performed on (b)(6) 2020.According to the complainant, during preparation, the physician noticed that the first spyscope ds frequently appeared black screen and could not show the image before the procedure.A second spyscope ds was used, however, the problem still occurred.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.Block e1 (initial reporter address 1): (b)(6).Block h6 (device codes): problem code 3191 is being used to capture the reportable issue of aborted/cancelled procedure.Block h6 (conclusion codes): evaluation conclusion code 4316 is being used in lieu of an adequate conclusion code for device not returned.Block h10: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
Note: this report pertains to a spyglass ds controller and two spyscope ds used during the same procedure.It was reported to boston scientific corporation that a spyglass ds controller and two spyscope ds were used during an endoscopic retrograde cholangiopancreatography (ercp) procedure in the common bile duct performed on (b)(6), 2020.According to the complainant, during preparation, the physician noticed that the first spyscope ds frequently appeared black screen and could not show the image before the procedure.A second spyscope ds was used, however, the problem still occurred.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.***additional information (b)(6) 2020*** reportedly, the two spyscope ds ii have been used with a different spyglass ds controller and worked as intended since this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS
Common Device Name
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10655640
Report Number3005099803-2020-04392
Device Sequence Number14145094
Product Code FBN
UDI-Device Identifier08714729863236
UDI-Public08714729863236
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K142922
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00546600
Device Catalogue Number4660
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/10/2020
Supplement Date Received by Manufacturer10/13/2020
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-