This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.In response to fda report number: mw5091823.A review of the device history record has been completed.No deviations or non-conformances noted.Further investigation summary: with the information collected during the investigation, there is enough evidence to support that device serial number (b)(4) was manufactured under controlled conditions in accordance with allergan procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run (b)(4) is not related to any additional complaint infection record reported as of today.During the trend review of all infection complaints for gel breast implants for the period of april 2018 through march 2020, was noted an outlier in sep-18.An additional analysis was performed to determine any pattern or any similarities relation between pr¿s involved.It was found that one primary packaging operator was involved in more than one occasion, however the person was trained in the corresponding procedure.Also, calibrations, maintenance and validations of the equipment involved in more than one occasion were reviewed and it was determined that they were performed on time and met specifications.In addition, all the records involved in this month were reviewed and no issues were found associated to either event.Given the fact that the cause of the infection cannot be specifically associated to manufacturing process, and there is not an adverse trend for this type of event no corrective action is deemed necessary at this time.The event of "infection" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: patient reported "in have had them removed and replaced numerous times due to what we thought was infection.".
|