Model Number 003500-0050-668
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
During the processing of this complaint, attempts were made to obtain complete event, patient, and device information.At this time it is unknown if the lens is being sent back thus, a proper device analysis could not be completed.Additionally, it was stated by the customer that there was nothing wrong with the device itself.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lenses that would have contributed to the nature of this complaint.Additionally, lenses are 100 % inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operation procedures and inspections and met all of the criteria for release.Failed injection/bleeding issues are known to be caused by numerous factors including, but not limited to: loading strategy, lens placement technique, poor handling during folding and inserting, patient pathology.The ifu instructs surgeons to carefully assess the patient's condition intraoperatively to determine if any of there are any complications present that maybe contraindicated.This assessment should be used to define a risk/benefit ratio for the purpose of determining if implantation of a zeiss iol is appropriate.This safety information is provided to the user as a notification of residual risk.We have reviewed the information provided and at this time there is no indication of a product quality issue with respect to the manufacturing, design, or labeling of the device.It appears that the reported issue is linked to user error.However, the need to explant the lens is considered need for medical intervention in order to prevent and avoid permanent impairment or damage to a body structure.
 
Event or Problem Description
It was reported that after injecting a ct lucia 602 the patient started bleeding the doctor the decided to remove the lens.It was stated that there was nothing wrong with the lens.No clinically significant delay in procedure reported.No additional information reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CT LUCIA 602
Common Device Name
CT LUCIA 602
Manufacturer (Section D)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s vintage ave. bld. a
ontario CA 91761
MDR Report Key10656164
Report Number3010126268-2020-00009
Device Sequence Number11645066
Product Code HQL
UDI-Device Identifier00843045100898
UDI-Public(01)00843045100898(17)241231
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P100016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/11/2020
Device Model Number003500-0050-668
Device Catalogue Number003500-0050-668
Device Lot Number3S20000430052
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-