Model Number 996081
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the balloon was burst.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be "marker bands damage balloon".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "1.Fill the inflation device (eagle¿ inflation device) with sterile, liquid media.2.Attach the inflation device to the balloon lumen.3.Open the stopcock and inflate the balloon.4.Once dilation has been attained, advance the sheath over the balloon.Warning: do not use air or any gaseous substances as a balloon inflation media, always use sterile liquid media.Note: the use of an inflation device with a pressure gauge is highly recommended to make sure adequate pressure is applied and the maximum limits of the balloon are not exceeded.Note: slight reduction in balloon pressure may be required if resistance is encountered when advancing the sheath over the balloon.Caution: do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended rbp".The user guide currently contains instruction that would assist in preventing potential user related causes of reported issue." the device was not returned.
 
Event or Problem Description
It was reported that the balloon was burst.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD X-FORCE NEPHROSTOMY BALLOON DILATION CATHETER
Common Device Name
DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10656446
Report Number1018233-2020-20031
Device Sequence Number2792920
Product Code LJE
UDI-Device Identifier10801741127653
UDI-Public(01)10801741127653
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063632
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2022
Device Model Number996081
Device Catalogue Number996081
Device Lot NumberBMDRFM29
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/16/2020
Supplement Date Received by Manufacturer05/03/2021
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-