• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RAD-8, VERTICAL; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RAD-8, VERTICAL; OXIMETER Back to Search Results
Model Number 22045
Medical Device Problem Codes Incorrect Measurement (1383); Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
 
Event or Problem Description
The customer reported inaccurate values and signal loss.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
 
Event or Problem Description
The customer reported inaccurate values and signal loss.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8, VERTICAL
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key10658124
Report Number3011353843-2020-00158
Device Sequence Number19619339
Product Code DQA
UDI-Device Identifier00843997005630
UDI-Public00843997005630
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K053269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number22045
Device Catalogue Number9191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/18/2020
Supplement Date Received by Manufacturer11/06/2020
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-