U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR WORLDWIDE LLC MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR WORLDWIDE LLC MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SMPX685
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: left breast infection.(b)(4).
 
Event or Problem Description
It was reported that a female patient who underwent a breast reconstruction procedure with a mentor memorygel breast implant 685cc gel breast prosthesis developed an infection in her left breast.As a result, the patient underwent explantation with no replacement device on (b)(6) 2020.
 
Additional Manufacturer Narrative
On 13-nov-2020, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced infection.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Infection complaint information is consistently analyzed and monitored by mentor quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2020, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR WORLDWIDE LLC
33 technology drive
irvine 92618
MDR Report Key10658394
Report Number1645337-2020-12844
Device Sequence Number3219871
Product Code FTR
UDI-Device Identifier00081317025160
UDI-Public00081317025160
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/25/2023
Device Model NumberSMPX685
Device Catalogue NumberSMPX685
Device Lot Number7646578
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Initial Date Received by Manufacturer 09/15/2020
Supplement Date Received by Manufacturer10/19/2020
11/13/2020
Initial Report FDA Received Date10/09/2020
Supplement Report FDA Received Date11/09/2020
11/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-