| Model Number |
07P9920 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A followup report will be submitted when the evaluation is complete.No patient information is available.
|
| |
|
Event or Problem Description
|
|
The account generated false elevated alinity c creatinine of 2.7 mg/dl on sid (b)(6) that repeated 0.66 mg/dl.The sample was repeated on another alinity analyzer with normal results.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there is normal complaint activity for lot 55012un19.Trending review determined no adverse trend for falsely elevated results for the product.Return testing was not completed as returns were not available.A review of the product labeling concluded that the issue is sufficiently addressed.Review in the corrective and preventive actions system did not identify any non-conformances or deviations related to the to the complaint list number, lot number and complaint issue.No product deficiency was identified for the alinity c creatinine reagent lot 55012un19.Section d.4.Catalog# updated from 07p99-20 to 07p99-30.
|
| |
|