PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20E
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2020
Event Type  malfunction  
Manufacturer Narrative
Physio-control evaluated the customer¿s device and verified the reported issue.Physio replaced the therapy pcb assembly and then observed proper device operation through functional and performance testing and the device was returned to the customer for use.Physio evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was due to integrated circuit, designator ic33 was electrically shorted.
 
Event Description
The customer contacted physio-control to report a non-critical issue with their device.Upon evaluation of the customer¿s device, physio-control observed that the device output was a monophasic shock with the negative portion of the biphasic waveform missing.The energy output delivered was reduced by 20% from the selected energy level.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key10659588
MDR Text Key211096212
Report Number0003015876-2020-01484
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873887707
UDI-Public00883873887707
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20E
Device Catalogue Number99507-000102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2020
Date Manufacturer Received09/25/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-