| Model Number |
FDS45030 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Stenosis (2263); Ischemia Stroke (4418); Embolism/Embolus (4438)
|
| Date of Event |
09/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject flow diverter was deployed across the neck of the unruptured posterior communicating artery aneurysm and there was narrowing at the mid-section of the subject device following deployment.Physician determined that the patient had underlying intracranial atherosclerotic disease in this portion of the internal carotid artery.During procedure, the physician tried to treat the device narrowing using a balloon but was not successful.Patient had embolic showers and some stroke like symptoms, six hours post implantation of the subject flow diverter.Unknown medication was given to the patient.Patient condition was stable.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device dfu.As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Based upon medical review, the harm observed in the complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted with the directions for use, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject flow diverter was deployed across the neck of the unruptured posterior communicating artery aneurysm and there was narrowing at the mid-section of the subject device following deployment.Physician determined that the patient had underlying intracranial atherosclerotic disease in this portion of the internal carotid artery.During procedure, the physician tried to treat the device narrowing using a balloon but was not successful.Patient had embolic showers and some stroke like symptoms, six hours post implantation of the subject flow diverter.Unknown medication was given to the patient.Patient condition was stable.No other information is available.
|
| |
|