• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT0206
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed, and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2020-01708.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the middle meningeal artery using penumbra smart coils (smart coils), a non-penumbra microcatheter, and non-penumbra catheter.It was reported that the patient's anatomy was tortuous.During the procedure, the physician implanted a smart coil in the target vessel using the microcatheter.While advancing the second smart coil in the introducer sheath, the smart coil did not advance smoothly and became stuck at the distal end of the introducer sheath.Subsequently, the smart coil was removed.The physician continued with the procedure, and implanted another smart coil.While advancing the fourth smart coil in the introducer sheath, the smart coil advanced a little before becoming stuck.Therefore, the smart coil was removed.The procedure was completed using another three smart coils, and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10659917
Report Number3005168196-2020-01707
Device Sequence Number14340664
Product Code HCG
UDI-Device Identifier00814548016030
UDI-Public00814548016030
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/10/2023
Device Model Number400SMTXSFT0206
Device Catalogue Number400SMTXSFT0206
Device Lot NumberF81388
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2020
Initial Report FDA Received Date10/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-