| Model Number |
68LP-480 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b.5., d.7.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side deflation.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: one opening curved on the posterior, crease fold, wear abrasion, yellow particles in the shell, and observed void >1 mm in non-textured, valve-to shell-bond area.Leak test and microscopic analysis was performed which identified: one opening crease sharp on the posterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: one crease sharp opening on the posterior assessed as fold flaw opening.Additional, changed, and/or corrected data: d.9., g.2., h.3., h.6.
|
| |
|