The reported field event of bpa was verified in the lab.However, further investigation showed the bpa was a false positive.Prior to the event, device had a reset, which cleared the device¿s usage data.The bpa algorithm uses device usage history data for calculations, and since that data got cleared after reset, a false bpa was triggered.This is normal and expected behavior when the data is cleared, and the device had been implanted for more than 6 years.Interrogation of the device revealed the device was above eri when received.No other anomaly was detected.The reported event or reset could not be confirmed.The device was not in backup operating mode upon receipt, additionally, a device image was not saved at the time of reset; the cause of reset could not be determined.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|