Model Number MX-410520
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Further investigation results: with the information collected during the investigation, there was enough evidence to support that the device was manufactured under controlled conditions in accordance with allergan procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run was not related to any additional complaint infection record reported at time of investigation.During the trend review of all infection complaints for gel breast implants for the period of jun/2018 through may/2020, two outliers were in sep/2018 and apr/2019.An additional analysis was performed to determine any pattern or any similarities relation between records involved.It was found that a primary packaging operator was involved in more than one occasion in this process, however the person was trained in the corresponding procedure.Also, it was found that a sealer and a sterilizer were involved in more than one occasion on those processes, however calibrations, maintenance, and the validation processes of these equipment involved were reviewed and it was determined that they were performed on time and met specifications.In addition, all the records involved in the current month were reviewed and no issues were found associated to either event.Given the fact that the cause of the infection cannot be specifically associated to manufacturing process, and there is not an adverse trend for this type of event, no corrective action is deemed necessary at time of investigation.The event of "infection" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection (unknown onset).
 
Event or Problem Description
Healthcare professional reported right side "patient's concern with the device".Additionally, patient reported right side infection.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key10661129
Report Number9617229-2020-12240
Device Sequence Number14429015
Product Code FTR
UDI-Device Identifier10888628003767
UDI-Public10888628003767
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/08/2021
Device Model NumberMX-410520
Device Catalogue NumberMX-410520
Device Lot Number2959214
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2020
Initial Report FDA Received Date10/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-