• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE Back to Search Results
Model Number M00561241
Medical Device Problem Codes Retraction Problem (1536); Failure to Cut (2587)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 09/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).Investigation analysis: a captivator ii-20mm round stiff snare was received for analysis.Visual evaluation revealed that the device did not have any defective condition.Functional evaluation of the returned device found that the device was tested in the 10-inch loop fixture and it was able to extend completely and retract without issues.Continuity test was also performed and the device passed the continuity test since the device's electrical resistance was 6.9 ohms, indicating a proper connection (the electrical resistance shall be less than 20 ohms per non-rotatable snares electrical resistance).A risk review of the captivator - snares was completed using the snare family global design fmea, bsc, bv and confirmed that the events of "snare- loop-failure to cut", "retraction problem" and "burn(s) " were defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications, however there is no control how the devices are handled/manipulated in the field.Based on the information available and the analysis performed, the most probable root cause for this problem is no problem detected since the device complaint or problem cannot be confirmed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator ii-20mm round stiff snare was used during a polypectomy procedure performed on (b)(6) 2020.According to the complainant, during the procedure and inside the patient, the wire could not get into the sheath and the device was unable to cut the target polyp.When the snare loop was slightly out, it was able to strangulate the polyp, so it was burnt.Reportedly, the snare was securely attached to the active cord and no visible issue was noted with the cautery pin.The procedure was completed with this device.The device caused a burn to the patient.The patient's condition following the procedure was reported to be no problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR II
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10661761
Report Number3005099803-2020-04522
Device Sequence Number4283803
Product Code FDI
UDI-Device Identifier08714729855927
UDI-Public08714729855927
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2023
Device Model NumberM00561241
Device Catalogue Number6124
Device Lot Number0025100885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/17/2020
Initial Report FDA Received Date10/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-