| Model Number |
222209 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.Udi:(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep via phone that during a rotator cuff repair the tip of the 5.5mm healix peek anchor w/orthocord broke upon insertion.Another device was used to complete the procedure.No patient consequences, or surgical delay reported.The device is available to be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during a rotator cuff repair the tip of the 5.5mm healix peek anchor w/orthocord broke upon insertion.The complaint was received and evaluated.Visual observation revelated that the anchor was broken from the distal part but is held in place by suture.Therefore, this complaint can be confirmed.Also, the suture card is still intact.There was no damage in the shaft.Finally, it was found tissue and blood debris along the anchor and the rest of the device.The possible root cause can be associated with off axis during insertion.However, it cannot be conclusively affirmed.A manufacturing record evaluation was performed for the finished device lot number: 4l35270, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|