Model Number 222209
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.Udi:(b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that during a rotator cuff repair the tip of the 5.5mm healix peek anchor w/orthocord broke upon insertion.Another device was used to complete the procedure.No patient consequences, or surgical delay reported.The device is available to be returned for evaluation.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during a rotator cuff repair the tip of the 5.5mm healix peek anchor w/orthocord broke upon insertion.The complaint was received and evaluated.Visual observation revelated that the anchor was broken from the distal part but is held in place by suture.Therefore, this complaint can be confirmed.Also, the suture card is still intact.There was no damage in the shaft.Finally, it was found tissue and blood debris along the anchor and the rest of the device.The possible root cause can be associated with off axis during insertion.However, it cannot be conclusively affirmed.A manufacturing record evaluation was performed for the finished device lot number: 4l35270, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX PEEK ANCH.W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10662201
Report Number1221934-2020-02950
Device Sequence Number17057818
Product Code MAI
UDI-Device Identifier10886705007578
UDI-Public10886705007578
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number222209
Device Catalogue Number222209
Device Lot Number4L35270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2020
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer10/14/2020
10/16/2020
Initial Report FDA Received Date10/12/2020
Supplement Report FDA Received Date10/15/2020
10/19/2020
Patient Sequence Number1
-
-