COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 8886848882
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic cholecystectomy, while the surgeon was trying to clip the bile duct, the clips were not able to be loaded properly and were not able to disengage from the cartridge.A new device was used to resolve the issue and complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d8, d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clips did not load properly into the jaws as expected.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key10662234
Report Number1219930-2020-04209
Device Sequence Number19801095
Product Code GCJ
UDI-Device Identifier10884521057678
UDI-Public10884521057678
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K925602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number8886848882
Device Catalogue Number8886848882
Device Lot NumberN8D0937X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer11/13/2020
Initial Report FDA Received Date10/12/2020
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-