• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1132
Device Problems Overheating of Device (1437); Inappropriate/Inadequate Shock/Stimulation (1574); Delayed Charge Time (2586); Charging Problem (2892)
Patient Problems Skin Irritation (2076); Discomfort (2330); No Code Available (3191)
Event Date 07/01/2020
Event Type  Injury  
Manufacturer Narrative
Exact date unknown, event occurred in (b)(6) 2020.
 
Event Description
It was reported that the patients ipg was not holding a charge and was taking longer to charge.It was also reported that during charging sessions, the ipg site would get warm and the patient felt shocking sensations.The ipg pocket was also irritated and sensitive to the touch.Database analysis revealed no anomalies.The patient underwent a revision procedure wherein the ipg was replaced and relocated.The patient was doing well postoperatively.The explanted ipg will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key10665313
MDR Text Key211017806
Report Number3006630150-2020-04803
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729821526
UDI-Public08714729821526
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2018
Device Model NumberSC-1132
Device Catalogue NumberSC-1132
Device Lot Number18934516
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/22/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
-
-