• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR101209
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2020
Event Type  malfunction  
Event Description
It was reported that during preventive maintenance, the anspach drills were overheating during the drill test. No patient was involved. No other complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANSPACH EMAX 2 PLUS HAND PIECE - ROHS
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
richard confer
2828 liberty ave
suite 100
pittsburgh, PA 15222
7634524980
MDR Report Key10667145
MDR Text Key211026836
Report Number3010266064-2020-01846
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/15/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberPFSR101209
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-